Plasmid concentration is finally measured using the Nanodrop 1000 photospectrometer. It is controlled by the "ND-1000" software (icon on the desktop).
Plasmid concentration is finally measured using the Nanodrop 1000 photospectrometer. It is controlled by the "ND-1000" software (icon on the desktop).
 +
 +
* '''Load nucleic acid program''': if necessary, click "exit" to return to the main menu, then "Nucleic Acid". Initialise as instructed, by depositing a ul drop of water on the sensor.
* '''Calibrate the spectrometer''': deposit a 1 ul drop of buffer solution onto the sensor, and click the "blank" button.
* '''Calibrate the spectrometer''': deposit a 1 ul drop of buffer solution onto the sensor, and click the "blank" button.
* '''Measure plamsid concentration''': deposit 1 ul of plasmid solution onto the sensor, and measure with the "measure" button. The reported plasmid concentration is considered valid if the "260/280" ratio is larger than 1.80.  
* '''Measure plamsid concentration''': deposit 1 ul of plasmid solution onto the sensor, and measure with the "measure" button. The reported plasmid concentration is considered valid if the "260/280" ratio is larger than 1.80.  
{{:Team:EPF-Lausanne/Templates/Footer}}