Reporter

What is a Reporter?

Being able to detect a particle in laboratory conditions can vary drastically from conditions in situ. After the presence of naphthenic acids is confirmed by our promoter, we will require a reporter system for our bacteria to report accurately and decisively. Over the course of the project, Team Calgary considered three different reporter systems - colorimetric, fluorescent, and electrochemical.

Why Electrochemical?

Figure 1 Tailings pond water is often murky and turbid, with an unknown composition of pollutants within.

Oilsands tailings ponds samples are difficult to work with for several different reasons. First of all, tailings ponds samples are murky, and can have greatly varying compositions from pond to pond. It would be difficult to obtain accurate colorimetric data from these samples, and it would be similarly difficult to readily observe fluorescence. Because an electrochemical response does not rely on being able to see anything, Team Calgary determined the electrochemical reporter to be ideal for this purpose.

Electrochemical reporters offer several advantages over colorimetric and flourometric reporters: the signal is robust enough to not be altered by the presence of pollutants or sample turbidity; the oxidation potential of our analyte is distinct and can be readily distinguished between our desired signal and background noise; and the data is immediately available for analysis.

A Novel Application for the lacZ Gene

Figure 2 Schematic diagram for the conversion of CPGR to CPR and Galactose. CPR can be oxidized at a specific voltage to produce an electrochemical output.
We decided to use the lacZ gene in our electrochemical reporter. Although traditionally, the lacZ gene (BBa_I732005) is used as a reporter for colorimetric assays (cleaving X-gal to create a blue pigment), the gene product ß-galactosidase, is capable of cleaving a variety of substrates. It has been shown by Biran et al. that ß-galactosidase can cleave p-aminophenyl-ß-D-galactopyranoside (PAPG), producing p-aminophenol (PAP). PAP can be oxidized at an electrode enabling detection of an electrochemical signal. It’s been shown that chlorophenolred-ß-D-galactopyranoside (CPRG) can also be cleaved by beta-galactosidase, producing chlorophenol red (CPR) and galactose. This substrate produces both a distinct color change (yellow to purple), and an electrochemical signal.

Building our System

One of the main reasons we selected a primarily electrochemical response for our system was that the response can be detected and processed digitally, without the need for subjective qualitative observations. The CPR analyte has an oxidation potential that is distinct from the tailings pond environment making it distinguishable from the background.

The voltammetric technique we used to detect chlorophenol red was cyclic voltammetry - the input voltage is shown below.

Figure 3 Applied potential across the cell during cyclic voltammetry.

As voltage slowly increases and decreases, each species present in solution will undergo a redox reaction at its respective oxidation potential. Because we know the oxidation potential of our analyte, chlorophenol red (CPR), we can ignore any background noise or pollution using this method.

A three-electrode setup is required to perform cyclic voltammetry. The electrical current passes between the working and counter electrodes - and the potential difference between the third electrode and counter electrode is controlled. This setup allows us to take voltage and current readings that do not interfere with each other.

Figure 4 A schematic for a potentiostat suitable for performing cyclic voltammetry. Source: Gopinath and Russell, 2005.

In order to perform cyclic voltammetry, specialized equipment is usually required. Because such equipment is usually unavailable in biology labs, we built our own potentiostat - a piece of equipment that outputs a known voltage and measures a corresponding current across two electrodes. Although we used a commercially available potentiostat to conduct our experiments, we decided to build our own to demonstrate that a cheap, reliable set-up is feasible.

Figure 4 A potentiostat setup capable of taking cyclic voltammetry measurements.
For preliminary data, click here.
For characterization data, click here.