Team:Calgary/Project/Promoter/Final Data

From 2011.igem.org

Revision as of 23:27, 28 September 2011 by Fwchung (Talk | contribs)


Electrochemical lacZ Characterization

Introduction

In order to be able to utilize the electrochemical detection system, various response levels must be correlated to cell concentration. Because our output signal is the product of a reaction sequence, it would not be sufficient to only characterize the performance of one particular part operating alone. This novel approach to characterizing lacZ through the oxidation of chlorophenol red was done in conditions allowing all the necessary reactions to occur simultaneously - just as if used in a field-ready biosensor.

The final results of our data processing is shown in Figure 1 below. The calculations done to arrive at this final curve follow.

Figure 1 Maximum rate of current change per voltage versus cell concentration for a standard electrochemical cell system.

By repeating this procedure for the various tested cell levels, we can construct one of our standard curves - the maximum rate of current change versus number of cells.

Appendix

In order to construct our standard curve for current response vs. number of cells, several experiments were conducted with varying numbers of cells, with a constant amount of CPRG (100uL at 0.5M), NaH2PO4 buffer (1mL at 0.5M), and nanopure water (up to 20mL). The cell was left for 5 minutes before the experiment was run.

Figure 2 below shows experimental data for a cyclic voltammetry experiment done with 7.5*105 cells.

Figure 2 Current plotted against voltage during a cyclic voltammetry run.

We observe the characteristic dip indicative of CPR oxidation at approximately -0.71V, showing the presence of CPR in solution. Next, the time derivative of this curve will be taken, shown below.

Figure 3 Rate of current change versus input voltage for a cyclic voltammetry assay.

Using Figure 3, we determine the maximum rate of change of current to occur at -0.755V, with a maximum of 258.45uA/V.

By repeating this procedure for the various tested cell levels, we can construct one of our standard curves - the maximum rate of current change versus number of cells.

Figure 4 Maximum rate of current change per voltage versus the number of cells.

Finally, by varying the amount of CPR in solution, and running cyclic voltammetry experiments in a simple system of only CPR, buffer, and nanopure water, we can construct a standard curve of CPR concentration versus maximum rate of current change. This graph is shown below.

Figure 5 Maximum rate of current change versus concentration of CPR in our electrochemical cell.

Currently, work is being done to refine our ability to differentiate between varying levels of CPR concentration.