Team:NCTU Formosa/protocol G
From 2011.igem.org
(Difference between revisions)
Line 28: | Line 28: | ||
#globalWrapper{ | #globalWrapper{ | ||
- | background-color: #565659; | + | //background-color: #565659; |
+ | background-image:url(https://static.igem.org/mediawiki/2011/a/ae/Wallpaper2.jpg); | ||
} | } | ||
Revision as of 16:17, 4 October 2011
Protocols
Gas chromatography (GC) Protocol
Method:Culture samples were centrifuged (12500×g, 25°C, 5 min), and the supernatants were analyzed for isobutanol. Isobutanol products were analyzed by GC equipped with flame ionization detector(FID) and a 50-m Ultra-2 column (Hewlett Packard, USA). The separation of alcohol compounds was carried out with a EQUity capirally column (SUPELCO, 60 m; 0.32 mm inside diameter; 1.0 μm film thickness). Filtrated samples were injected in split injection mode (1:10 split ratio).Nitrogen (constant flow 1 mL/min) was used as a carrier gas. The temperature of the injector was 150 °C. The oven program was as follows: 40°C ramp to 100 °C at 20 °C/min, 100 °C for 1 min, ramp to 130 °C at 15 °C/min, 130°C for 4 min.