Team:Northwestern/Results/Summary

From 2011.igem.org

Revision as of 00:40, 29 October 2011 by Helenmelon (Talk | contribs)

RETURN TO IGEM 2010



Overview


We characterized more than two dozen parts over the course of this project. Here, we focus our discussion on the biosensor constructs that best illustrate our successful proof-of-concept experiments. These parts are [lasP+RBS30+GFP, CP+RBS30+lasR], [rhlP+RBS30+GFP, CP+RBS30+rhlR], and [GR(S)+RBS34+GFP, CP+RBS34+RhlR].


PAI-1 Biosensor: Binary Detection System


In order to evaluate the suitability of our biosensor constructs for detecting P. aeruginosa, we conducted a series of dose-response studies for characterization purposes. We also analyzed this data to determine the transfer function and dynamic range of each biosensor system.


Figure 1: Dose response of the PAI-1 biosensor system (LasP+RBS30+GFP and CP+RBS30+lasR). Immediately before the assay, cells were diluted to ensure that they were growing in the exponential phase for the experiment. Autoinducer PAI-1 was added at the concentrations indicated, and GFP fluorescence was quantified using an incubated, shaking plate reader. In this plot, fluorescence was normalized to the culture OD to control for cell growth. Samples were run in quadruplicates with standard deviation indicated (error bars = SD; n=4).
fig1


Our first observation was that the construct [lasP+RBS30+GFP, CP+RBS30+lasR] appears to be well-suited as a binary sensor and can be used to simply determine whether or not P. aeruginosa is present in a sample. The construct follows the same general trend of fluorescence at varying autoinducer concentrations, except for the negative control in which no autoinducer was present. There is a significant amount of overlap in the error bars in Figure 1, so T-tests were conducted in order to evaluate the statistical significance of this observed trend, as detailed below in Table 1.


Table 1: Statistical analysis of the PAI-1 biosensor response (lasP+RBS30+GFP, CP+RBS30+lasR). The null hypothesis is that there is no statistical difference between the means of the compared samples. Green cells indicate rejection of the null hypothesis (p<0.05), while blue cells indicate failure to reject (p>0.05). (A) Data from each of the initial segments of the curves (region before any fluorescence is observed ~30min) was compared between varying autoinducer concentrations. (B) Data from the initial segment of each curve (before any fluorescence is observed) was compared with the final steady state fluorescence of that curve from the same sample (last 10 data points). (C) Data from each of the final steady state segments of the curves was compared between varying autoinducer concentrations (last 10 data points).
fig1


In the first 30 minutes after autoinducer dosages were given, only the sample that was given 100μM of PAI-1 exhibited a response that was significantly different from the other samples (Table 1A). Table 1b tests the activation of each construct. This is accomplished by comparing the data from the initial segment of the curves (before any fluorescence is observed) with the final steady state fluorescence (last 10 data points). Our data also indicated that fluorescence per OD changed significantly over time in each sample (Table 1B).

Table 1C compares the data from each of the final steady state segments of the curves with the other final steady state segments (last 10 data points). The negative control exhibited significantly different fluorescence readings per OD. This is because the cells showed relatively steady total fluorescence coupled with rapid of cell growth, resulting in a decrease in fluorescence per OD, as shown in Figure 2 below. All the samples showed similar fluorescence readings except for the 0.1μM, to some extent the 0.5μM, and of course 0μM PAI-1 negative control.


Figure 2: Overall response of the PAI-1 biosensor system (lasP+RBS30+GFP, CP+RBS30+lasR). Here, the data from Figure 1 are presented without normalization, showing total fluorescence (left) and total OD of the cultures.
fig1


In order to further characterize this biosensor, we next plotted the steady state fluorescence (per OD) vs. autoinducer concentration to determine the input-output transfer function (Figure 3). As indicated by the analysis and discussion above, in this “binary” biosensor, fluorescence per OD stayed relatively constant (within about 10% of the mean fluorescence per OD) for all samples treated with PAI-1 autoinducer, and all samples exhibited significantly different behavior in contrast to the negative control sample.


Figure 3: Input-output transfer function for PAI-1 biosensor (lasP+RBS30+GFP, CP+RBS30+lasR). Steady-state responses were calculated from the data in Figure 1 and plotted against the input concentration of PAI-1 autoinducer.
fig1


PAI-2 Biosensor 1: Concentration Detection System


In contrast to the binary detection system, the construct [rhlP+RBS30+GFP, CP+RBS34+rhlR] is well suited for determining the amount of P. Aeruginosa present in a sample because it appears to be able to detect and discriminate between varying autoinducer concentrations. Figure 4 below details the fluorescence per OD observed upon the induction of the system at multiple autoinducer concentrations. Unlike the binary detection system, the fluorescence per OD of each curves is distinctive.


Figure 4: Dose response of the PAI-2 Biosensor System 1 (rhlP+RBS30+GFP, CP+RBS34+rhlR). This experiment was conducted as in Figure 1, and strains were exposed to 0 - 100 micromolar PAI-2. Fluorescence and OD were measured every 7.5 min (error bars = SD; n=4).
fig1


For the most part, error bars only overlapped when the autoinducer concentration exceeded 15μM, which is not a physiologically relevant value. Notably, autoinducer concentrations between 0μM and 5μM are clear and distinguishable. T-tests were conducted in order to statistically confirm the variance of each concentration curve, as detailed below in Table 2.


Table 2: Statistical analysis of the PAI-2 Biosensor 1 response (rhlP+RBS30+GFP, CP+RBS34+rhlR). The null hypothesis is that there is no statistical difference between the means of the compared samples. Green cells indicate rejection of the null hypothesis (p<0.05), while blue cells indicate failure to reject (p>0.05). (A) Data from each of the initial segments of the curves (region before any fluorescence is observed ~30min) was compared between varying autoinducer concentrations. (B) Data from the initial segment of each curve (before any fluorescence is observed) was compared with the final steady state fluorescence of that curve from the same sample (last 10 data points). (C) Data from each of the final steady state segments of the curves was compared between varying autoinducer concentrations (last 10 data points).
fig1


Table 2A, as before, compares the early time points, before GFP is induced (t=30 min). As we observed with the PAI-1 biosensor, only the 100μM sample significantly induced fluorescence in the PAI-2 Biosensor 1 within the first 30 minutes.


Table 2B, as before, shows promoter activation, by comparing the data from the initial segment of the curves (before any fluorescence is observed) with the final steady state fluorescence (last 10 data points). The system exhibited behavior that is similar to the behavior of the binary sensor, as each construct changes a statistically significant amount when in the presence of autoinducers. Surprisingly, the 0μM autoinducer sample also has a statistically significant change. However, the induced constructs produced fluorescence readings that were orders of magnitude higher than the negative control, so any cell bias introduced by an OD irregularity is insignificant and can be ignored.


Table 2C compares the data from each of the final steady state segments of the curves with the other final steady state segments (last 10 data points). The data shows that the final steady-state fluorescence of nearly every sample is distinct in comparison to the final-steady state fluorescence of samples that were administered a different autoinducer concentration. The two exceptions are the 7.5μM-10μM, and the 15μM-20μM autoinducer concentration samples. It is important to note that 10μM-15μM and 20μM-50μM concentration samples have statistically significant responses. In fact, the high degree of discrimination between relative autoinducer concentrations strongly qualifies this construct as a useful concentration-based detection method. The steady state fluorescence per OD is presented in Figure 5. The logarithmic regression fits the data quite well, and proves that the construct can be used in a biosensor application.


Figure 5: Input-output transfer function for PAI-2 Biosensor 1 (rhlP+RBS30+GFP, CP+RBS34+rhlR). Steady-state responses were calculated from the data in Figure 4 and plotted against the input concentration of PAI-2 autoinducer.
fig1


PAI-2 Biosensor 2: Newly Isolated Genomic Promoter Construct (Novel Entry)


In addition to designing plasmid systems using existing biobricks from the registry, we also engineered a plasmid that contains an inducible promoter that was isolated from the genome of P. aeruginosa. Out of all the parts that we derived from the P. aeruginosa genome, the construct [GR(S)+RBS34+GFP, CP+RBS34+RhlR] was most suitable in autoinducer detection. Figure 6 shows our GR(S) promoter is sensitive to a range of autoinducer concentrations and requires an average of two hours for reach full induction.


Figure 6: Dose response of the PAI-2 Biosensor System 2 (GR(S)+RBS34+GFP, CP+RBS34+RhlR). This experiment was conducted in a similar fashion as the experiment shown in Figure 1, and strains were exposed to 0 - 100 micromolar PAI-2. Fluorescence and OD were measured every 7.5 min (error bars = SD; n=4).
fig1


Just as in the previous concentration-dependent detection system, the genomic promoter system also discriminates very well between the varying autoinducer concentrations. To confirm our results, we conducted T-tests to validate the observed significance between the curves. The T-tests conducted follow the same format as the prior ones described above and is detailed below in Table 3.


Table 3: Statistical analysis of the PAI-2 Biosensor 2 response (GR(S)+RBS34+GFP, CP+RBS34+RhlR). The null hypothesis is that there is no statistical difference between the means of the compared samples. Green cells indicate rejection of the null hypothesis (p<0.05), while blue cells indicate failure to reject (p>0.05). (A) Data from each of the initial segments of the curves (region before any fluorescence is observed ~30min) was compared between varying autoinducer concentrations. (B) Data from the initial segment of each curve (before any fluorescence is observed) was compared with the final steady state fluorescence of that curve from the same sample (last 10 data points). (C) Data from each of the final steady state segments of the curves was compared between varying autoinducer concentrations (last 10 data points).
fig1


Table 3A shows that the genomic promoter construct had better uniformity in early time points than our concentration detection system (discussed above). The initial data of almost every sample coincided with that of the negative control (0μM). Once again, the 100μM sample exhibited induced fluorescence quickly. Thus, virtually all the samples exhibit fluorescence similar to that of the negative control at this short time point.


In Table 3B, the data shows that similar to previous systems, each construct exhibited a statistically significant amount of fluorescence per OD in the presence of autoinducers. The only outlying T-test in this characterization is the 0.1μM-20μM comparison, which may indicate that the uninduced state of the 20μM autoinducer sample was close to the induced state of the 0.1μM autoinducer sample’s fluorescence.


Table 3C's data validates the claim that the fluorescence per OD of every single steady state curve is statistically different across varying autoinducer concentrations. This result is further highlighted by the steady state fluorescence per OD vs. autoinducer concentration in Figure 8. The low error and an extremely well fit logarithmic regression exhibited proves that the construct can not only works exactly the way it should, but is also the most suitable construct in a biosensor application.


Figure 7: Input-output transfer function for PAI-2 Biosensor 2 (GR(S)+RBS34+GFP, CP+RBS34+RhlR). Steady-state responses were calculated from the data in Figure 6 and plotted against the input concentration of PAI-2 autoinducer.
fig1


Model Fitting and Analysis


In the modelling section, the theoretical math behind the system was discussed. In order to fully understand our system, the mathematical model had to support the experimental data. Consequently, the sensitivity analysis was conducted (also in the modelling section) to gauge which factors influenced the system the most. A number of parameters were found to exert a significant amount of influence on the system. Each of these parameters was carefully adjusted until the theoretical graphs closely resembled those received from experimental testing as demonstrated below in Figure 8.


Figure 8: Curve fit to the experimental data using the math model
fig1


Upon administration of the autoinducer concentration, the intracellular concentration of the autoinducer increases dramatically due to its passive transport through the cell membrane. Similarly, the initial concentration of R-protein is quite high relative to the other biochemical species in the cell due to its constitutive production. However, the increase in intracellular autoinducer concentration facilitates an instantaneous drop in the (free) R-protein concentration. The R-protein/dimer concentration level increases almost simultaneously, inducing the R-protein/autoinducer induced promoter. The sudden rise in the dimer complex facilitates a sudden increase in expression of the reporter construct or GFP. The concentration of GFP continually increases due to the increasing concentration of the dimer complex.


Autoinducer concentrations continue to drop until the reversible binding of the R-protein, autoinducer and the dimer complex reach steady state. This approximately happens at 150 minutes. At this point, most the autoinducer is bound to the dimer complex, and the remaining amount is degrading. In a way, the dimer complex is protecting the autoinducer from degradation, which maintains steady r-protein and dimer complex levels. Meanwhile GFP is being generated due to the initial excessive amount of reporter expression. However, when the R-protein and dimer complex reach steady state, so does GFP expression. additionally, the model fits the fluorescence data in the above characterization graphs. If one were to closely observe the GFP curve, the concentration actually decreases slightly over an extended period of time, which is exactly what was observed experimentally.


Conclusion


On the whole, we demonstrated that our biosensor constructs were functional and are excellent candidates for a cell-based biosensor system for either PAI-1 or PAI-2. Both the Las and Rhl reporter constructs exhibited fluorescence that was statistically significant relative to the controls at most or all of the autoinducer concentrations we tested. Compared to the Las (PAI-1) construct, the Rhl (PAI-2) systems produced more fluorescence overall and output levels were more strongly dependent on the concentration of autoinducer used (potentially indicating a broader dynamic range for these biosensors). The Las construct produced very similar fluorescence values at almost every autoinducer concentration. Our results provide detailed characterizations of both the original registry Las & Rhl promoters on which our constructs are based, as well as the newly assembled biosensor constructs we generated. Our results also provide a solid basis for a synthetic cell-based “device” that can detect the presence of Pseudomonas Aeruginosa.


Our RFP constructs were not as successful. The results were inconsistent and the signals did not seem to be correlated to the presence of the autoinducers. Although we were not able to determine the cause of this problem, we have put the parts in the registry for future investigation.


In addition to working with the Las and Rhl dependent promoters already on the registry, we also characterized some new promoters isolated directly from the Pseudomonas genome. This resulted in another potential set of promoters for use in a Pseudomonas detection device. In particular, the Rhl-dependent promoter we extracted from the genome showed excellent sensitivity to PAI2.


For complete characterization information on all of our parts, please refer to our Biobricks page. Clicking on a part will take you to the Registry of Standard Biological Parts, where the complete analysis for that part is posted.


Refer here for our testing protocol.



Sponsor northwestern.jpg Sponsor weinberg.jpg Sponsor mccormick.jpg