Team:Cambridge/Protocols/Confocal Microscopy of Loligo Eye and Mantle Dermis Samples

From 2011.igem.org

(Difference between revisions)
(Confocal Microscopy)
Line 1: Line 1:
{{Template:Team:Cambridge/CAM_2011_TEMPLATE_HEAD}}
{{Template:Team:Cambridge/CAM_2011_TEMPLATE_HEAD}}
-
__NOTOC__
 
-
 
=Confocal Microscopy=
=Confocal Microscopy=

Revision as of 09:18, 19 September 2011

Loading...
OVERVIEW
home

Contents

Confocal Microscopy

Theory

Confocal microscopy uses a pinhole aperture to direct light to a small region of a sample. Illuminating one small point at a time reduces the scattering of light which would normally obscure an image. Light is collected via another pinhole. This second pinhole further enhances the image because it only permits the entry of light that is directly from the focal point. Due to the punctate nature of the imaging process, images must be reconstructed, in modern confocal microscopes a computer assembles the array of pixels into an on screen image. As described above we set the microscope to collect light reflected from the sample as we were searching for iridescence in the samples.

Practice

Preparing Squid Samples

For observations of squid reflectin, samples were dissected from fresh Loligo vulgaris using disposable scalpels and razor blades. These were mounted on glass microscope slides with PBS and cover slips.

Preparing Bacterial Samples

Microscopy

The confocal microscope was set to capture reflected light (the receiver wavelengths encompassed the incident laser light) and incident beams of 405nm, 488nm, 561nm and 633nm were used. The images were overlaid digitally to create a composite.

We are very grateful to Paul Grant and Fernan Federici for their assistance in optimising these settings.

Safety

The main dangers to be aware of are scalpels, glass slides and the laser within the microscope.

Disposable scalpels were used, which are easily disposed of in the yellow sharps bins in the labs. Broken glass should be reported to the lab manager and carefully brushed up and disposed of in the allocated glass bin. The laser is only dangerous when the filter block is being rotated, so the operator of the microscope should inform anyone when this is occuring and make sure no one is looking at the eye-pieces when the filter blocks are changed. Using the computer monitor for observation eliminates the risk of eye damage from the lasers.